51
IMPORTANT
•Wasthesampledeepredincolorandfreeofwateryuid?
•Wasthedroplargeenough(atleast0.3µL)?
•Didthereactionsitellwithenoughbloodallatonce?
Step 5:
The meter will count down and display result in 7 seconds.
CONSIDER YOUR RESULT. REPEAT ALTERNATE SITE
TEST:
•IFthebloodsampleappearedtobedilutedwithclearuid.
•IFyoudidnotvigorouslyrubthetestsite.
•IFtheblooddropwasnotlargeenoughtollthe
reaction site.
•IFthetestwasaccidentallymarkedasa“Control”result.
•IFyourresultwasnotconsistentwithhowyoufeel.
•IFmorethan20secondselapsedfromsamplecollectionto
measurement (evaporation of the blood sample may cause a test
result that is higher than the actual value).
IF NOT,
REPEAT TEST.
Filled
Not Filled
Filled
Not Filled
Any of the above situations can lead to an inaccurate test result.